DIRECTIONS FOR USE
Allow the test device, specimen, buffer, and/or controls to reach room temperature (15-30°C) prior to testing.
1. Bring the pouch to room temperature before opening. Remove the test device from the sealed pouch and use it as soon as possible.
2. Place the test device on a clean and level surface.
For Serum or Plasma Specimens: Using the provided 5uL disposable pipette,and transfer 1 drop of serum/plasma to the specimen well of the test device,then add 2 drops of buffer and start the timer.
For Whole Blood (Venipuncture/Fingerstick) Specimens: Using the provided 5uL disposable pipette,and transfer 2 drop of whole blood (approximately 20µL) to the specimen well of the test device,then add 2 drops of buffer and start the timer. Note: Specimens can also be applied using a micropipette.
3. Wait for the colored line(s) to appear.Read results at 10 minutes.Do not interpret the result after 15 minutes.